LOGOEEEC.JPG

DICIEMBRE 2017

CICLO 135

DAO:DROGAS DE ABUSO EN ORINA

REPORTE DE RESULTADOS CUALITATIVOS 

REACTIVO

1.00

INCIERTO

0.5

NO REACTIVO

0.01

NO SE PROCESÓ

LA MUESTRA DE ORINA ENVIADA EN EL CICLO ES UN CONTROL NEGATIVO  (Ratio < 0.9 )  PARA TODAS LAS DROGAS

ANFTM

BARB

BNZDP

CANNA

COCA

ETANOL

OPIAC

MEDIANA CONTROL CUANTITATIVO

122

8

2

9

23

8

82

CUTOFF

500

200

200

50

150

40

300

RATIO

0.2

0.0

0.0

0.2

0.2

0.2

0.3

       

DAO

D.U

pH

CREAT

ANFTM

BARB

BNZDP

CANNA

COCA

ETANOL

OPIAC

N

10

10

11

9

4

7

10

10

5

8

MAX

1.020

7.0

86.8

0.01

0.01

0.01

0.01

0.01

1.00

0.01

% REACTIVO

1.013

6.1

74.0

0%

0%

0%

0%

0%

18%

0%

MIN

1.010

5.5

50.0

0.01

0.01

0.01

0.01

0.01

0.01

0.01

D.U

pH

CREAT

ANFTM

BARB

BNZDP

CANNA

COCA

ETANOL

OPIAC

135:15/12/2017 03:26:07 p. m.,BECKMAN COULTER,5.00

1.010

6.0

86.8

0.0

0.0

0.0

0.0

0.0

0.0

0.0

135:22/12/2017 09:05:45 p. m.,BECKMAN COULTER,22.00

1.010

6.5

78.2

0.0

0.0

1.0

135:27/12/2017 02:12:45 p. m.,ARCHITECT DE ABBOT,10.00

1.010

6.0

83.9

0.0

0.0

0.0

0.0

0.0

0.0

0.0

135:20/12/2017 10:53:47 a. m.,BECKMAN COULTER,6.00

1.015

6.5

81.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

135:13/12/2017 12:08:39 p. m.,SIEMENS,18.00

1.011

6.0

75.6

0.0

0.0

0.0

135:28/12/2017 12:54:03 p. m.,ARCHITECT ci16200,18.00

1.015

6.0

75.0

0.0

0.0

0.0

0.0

0.0

135:22/12/2017 01:32:01 p. m.,INMUNOCROMATOGRAFICO,20.00

1.015

6.0

63.9

0.0

0.0

0.0

0.0

135:31/12/2017 02:56:37 p. m.,OPERADOR VISUAL,25.00

1.015

6

68.9

0.0

0.0

0.0

0.0

0.0

0.0

135:26/12/2017 10:52:15 a. m.,inmunocromatografia,26.00

1.010

6.0

65.6

0.0

0.0

0.0

0.0

0.0

135:23/12/2017 08:47:52 a. m.,ARCHITECT,12.00

71.7

0.0

135:15/12/2017 12:14:12 p. m.,INMUNOENSAYO,28.00

1.011

6.0

76.2

0.0

0.0

0.0

0.0

0.0

135:20/12/2017 10:16:31 a. m.,BECKMAN COULTER,5.00

1.020

6.0

83.6

0.0

0.0

0.0

0.0

0.0

0.0

135:19/12/2017 01:09:59 p. m.,SIEMENS,24.00

1.010

6.0

72.3

0.0

0.0

0.0

0.0

0.0

135:18/12/2017 12:55:40 p. m.,Instant View visual,22.90

1.015

7.0

50.0

0.0

0.0

0.0

0.0

0.0

0.0

135:22/12/2017 04:28:51 p. m.,Método Manual; Inmunocromatográfico,21.00

1.010

6.0

77.1

0.0

0.0

0.0

0.0

0.0

0.0

135:22/12/2017 03:49:23 p. m.,Architect C8000,4.00

1.011

6.2

76.5

0.0

0.0

0.0

0.0

0.0

135:19/12/2017 06:43:14 a. m.,Prueba de Casette,25.00

0.0

0.0

0.0

0.0

0.0

135:16/12/2017 12:36:31 p. m.,visual,20.00

0.0

0.0

135:22/12/2017 05:01:30 p. m.,MANUAL,20.00

0.0

0.0

0.0

0.0

135:22/12/2017 12:54:34 p. m.,DIAGMEX,27.00

1.015

6.0

0.0

0.0

0.0

0.0

135:22/12/2017 12:54:34 p. m.,DIAGMEX,27.00

1.015

6.0

0.0

0.0

0.0

0.0

135:29/12/2017 10:05:08 a. m.,Architect CI8200,20.00

1.011

6.5

72.7

0.0

0.0

0.0

135:19/12/2017 04:27:32 p. m.,Architect c16000,16.10

1.020

6.0

74.0

0.0

0.0

0.0

0.0

0.0

0.0

0.0

135:20/12/2017 01:00:06 p. m.,SIEMENS,12.00

1.010

6.0

73.9

0.0

0.0

0.0

0.0

0.0

135:18/12/2017 08:14:57 a. m.,SIEMENS,18.00

1.015

5.5

72.8

0.0

0.0

0.0

0.0

0.0

135:21/12/2017 11:01:30 a. m.,TIRAS REACTIVAS,25.00

0.0

0.0

0.0

0.0

0.0

0.0

1.- LAS PRUEBAS VISUALES EN TIRAS REACTIVAS AUNQUE SON ECONOMICAS Y RAPIDAS SON POCO CONFIABLES POR LO QUE LOS LABORATORIOS CLINICOS NO LAS DEBERIAN UTILIZAR.

TODOS LOS RESULTADOS EN PRINCIPIO DEBERIAN SER REALIZADOS CON SISTEMAS INMUNOMETRICOS CON SENSIBILIDAD Y ESPECIFICIDAD ESTABLECIDA DEJANDO LAS PRUEBAS

RAPIDAS PARA SITUACIONES DE URGENCIA EXCLUSIVAMENTE.

2.- CUANDO SE UTILIZAN CONTROLES INDEPENDIENTES ( TERCERA OPINION ) EN LAS PRUEBAS INMUNOMÈTRICAS SE PUEDE ENCONTRAR UN

EFECTO MATRIZ EN ELLOS, ESTE FENOMENO SE PUEDE DEBER A LA PRESENCIA DE MEDICAMENTOS, METABOLITOS U OTROS COMPUESTOS EN LOS MISMOS.

3.- EL EFECTO MATRIZ TAMBIEN PUEDE PRESENTARSE EN LAS MUESTRAS DE LOS PACIENTES POR ALTERACIONES EN SUS METABOLITOS ( PROTEINAS, LIPIDOS, ETC)

PERO SOBRE TODO CUANDO ADEMÀS DE LOS CAMBIOS PROPIOS DE LA ENFERMEDAD LOS PACIENTES ESTAN BAJO TRATAMIENTO CON MULTIPLES MEDICAMENTOS.

4.- LA CONFIABILIDAD DE LOS RESULTADOS DEPENDE DEL VALOR DE CORTE UTILIZADO LO CUAL DETERMINA DE MANERA SIGNIFICATIVA LA SENSIBILIDAD Y DE LA ESPECIFICIDAD DE LOS SISTEMAS ANALITICOS POR LO QUE

PARA HACER EL DIAGNOSTICO DEFINITIVO ES NECESARIO INFORMAR LA NORMA UTILIZADA ADEMÀS DE CONFIRMAR TODAS LAS PRUEBAS REACTIVAS UTILIZANDO METODOS MAS ESPECIFICOS ( HPLC-GC-EM )

 

BAYES.jpg

EL VALOR DE CORTE ( CUTOFF ) UTILIZADO EN LOS ESTADOS UNIDOS ES DIFERENTE AL DE LA UNION EUROPEA SOBRE TODO EN COCAINA Y OPIACEOS

LAS NORMAS ISO 151899 E ISO 17043 SON NORMAS EUROPEAS POR LO QUE EN EL EEEC QUALITAT  UTILIZAMOS EL CUTOFF DE EUROPA

DAO.USA.jpg

DAO.USA.jpg

DAO.2.jpg

METODOS.jpg

ISO 13528 STATISTICAL METHODS FOR USE IN PROFICIENCY TESTING BY INTERLABORATORY COMPARISON.2015

METODOS.jpg

Enter supporting content here